“Hook-like Effect” Causes False-negative Point-of-care Urine Pregnancy Testing in Emergency Patients

      Abstract

      Background

      Failure to detect pregnancy in the emergency department (ED) can have important consequences. Urine human chorionic gonadotropin (uhCG) point-of-care (POC) assays are valued for rapidly detecting early pregnancy with high sensitivity. However, under certain conditions, POC uhCG tests can fail to detect pregnancy.

      Objectives

      In investigating a series of late first-trimester false-negative pregnancy tests in our ED, a novel and distinct causative phenomenon was recently elucidated in our institution. We discuss uhCG POC tests, review our false-negative rate, and describe mechanisms for false negatives and potential remedies.

      Discussion

      The false-negative POC uhCG rate is very low, but in the setting of a large volume of tests, the numbers are worth consideration. In positive uhCG POC tests, free and fixed antibodies bind hCG to form a “sandwich”; hCG is present in several variant forms that change in their concentrations at different stages of pregnancy. When in excess, intact hCG can saturate the antibodies, preventing sandwich formation (hook effect phenomenon). Some assays may include an antibody that does not recognize certain variants present in later stages of pregnancy. When this variant is in excess, it can bind one antibody avidly and the other not at all, resulting in a false-negative test (hook-like phenomenon). In both situations, dilution is key to an accurate test.

      Conclusions

      Manufacturers should consider that uhCG tests are routinely used at many stages of pregnancy. Characterizing uhCG variants recognized by their tests and eliminating lot-to-lot variability may help improve uhCG test performance. Clinicians need to be aware of and familiarize themselves with the limitations of the specific type of uhCG POC tests used in their practice, recognizing that under certain circumstances, false-negative tests can occur.

      Keywords

      To read this article in full you will need to make a payment

      References

        • Brewer C.
        Failure to diagnose pregnancy as cause of late abortion [Letter].
        Lancet. 1977; 1: 46
        • Seymour J.A.
        Negligent failure to provide appropriate information in obstetrics and midwifery.
        in: Childbirth and the law. Oxford University Press, New York2000: 82-101
        • Morrison L.J.
        General approach to the pregnant patient.
        in: Marx J. Hochberger R. Walls R. 7th edn. Rosen's emergency medicine: concepts and clinical practice. Vol. 2. C.V. Mosby, St. Louis, MO2009: 2268-2278
        • Ferman G.J.
        • Suyama J.
        Point of care testing in the emergency department.
        J Emerg Med. 2002; 22: 393-404
        • Lazarenko G.C.
        • Dobson C.
        • Enokson R.
        • Brant R.
        Accuracy and speed of urine pregnancy tests done in the emergency department: a prospective study.
        CJEM. 2001; 3: 292-295
        • Gronowski A.M.
        • Cervinski M.A.
        • Stenman U.H.
        • Woodworth A.
        • Ashby L.
        • Scott M.G.
        False-negative results in point-of care qualitative human chorionic gonadotropin (hCG) devices due to excess hCGb core fragment.
        Clin Chem. 2009; 55: 1389-1394
        • Butler S.A.
        • Khanlian S.A.
        • Cole L.A.
        Detection of early pregnancy forms of human chorionic gonadotropin by home pregnancy test devices.
        Clin Chem. 2001; 47: 2131-2136
      1. Interview by Bob Barrett of Clinical Chemistry with Ann M. Gronowski on false negative results in qualitative hCG devices due to excess hCG beta core fragment; July 27, 2009 [podcast transcript]. Available at: http://www.aacc.org/events/podcast/Documents/072709Gronowski.pdf. Accessed September 14, 2009.

        • Fritz M.A.
        • Guo S.M.
        Doubling time of human chorionic gonadotropin (hCG) in early normal pregnancy: relationship to hCG concentration and gestational age.
        Fertil Steril. 1987; 47: 584-589
        • Wilcox A.J.
        • Baird D.D.
        • Dunson D.
        • McChesney R.
        • Weinberg C.R.
        Natural limits of pregnancy testing in relation to the expected menstrual period.
        JAMA. 2001; 286: 1759-1761
        • Cervinski M.A.
        • Lockwood C.M.
        • Ferguson A.M.
        • et al.
        Qualitative point-of-care and over-the-counter urine hCG devices differentially detect hCG variants of early pregnancy.
        Clin Chem Acta. 2009; 406: 81-85
        • Gronowski A.M.
        • Powers M.
        • Stenman U.H.
        • Ashby L.
        • Scott M.G.
        False-negative results from point-of-care qualitative human chorionic gonadotropin (hCG) devices caused by excess hCGbeta core fragment vary with device lot number.
        Clin Chem. 2009; 55: 1885-1886
        • Kato Y.
        • Braunstein G.D.
        Beta-core fragment is a major form of immunoreactive urinary chorionic gonadotropin in human pregnancy.
        J Clin Endocrinol Metab. 1988; 66: 1197-1201
        • Sturgeon C.M.
        • Berger P.
        • Bidart J.M.
        • et al.
        IFCC Working Group on hCG. Differences in recognition of the 1st WHO international reference reagents for hCG-related isoforms by diagnostic immunoassays for human chorionic gonadotropin.
        Clin Chem. 2009; 55: 1484-1491
        • Cole L.A.
        Immunoassay of human chorionic gonadotropin, its free subunits, and metabolites.
        Clin Chem. 1997; 43: 2233-2243