A Randomized Controlled Trial using iTClamp, Direct Pressure, and Balloon Catheter Tamponade to Control Neck Hemorrhage in a Perfused Human Cadaver Model

      Abstract

      Background

      Penetrating neck wounds are common in the civilian and military realms. Whether high or low velocity, they carry a substantial morbidity and mortality rate.

      Objectives

      We endeavored to ascertain whether the iTClamp is equivalent to direct manual pressure (DMP) and Foley catheter balloon tamponade (BCT).

      Methods

      Using a perfused cadaver, a 4.5-cm wound was made in Zone 2 of the neck with a 1-cm carotid arteriotomy. Each of the hemorrhage control modalities was randomized and then applied to the wound separately. Time to apply the device and fluid loss with and without neck motion was recorded.

      Results

      There was no significant difference between the fluid loss/no movement (p > 0.450) and fluid loss/movement (p > 0.215) between BCT and iTClamp. There was significantly more fluid lost with DMP than iTClamp with no movement (p > 0.000) and movement (p > 0.000). The iTClamp was also significantly faster to apply than the Foley (p > 0.000).

      Conclusions

      The iTClamp and BCT were associated with significantly less fluid loss than DMP in a perfused cadaver model. The iTClamp required significantly less time to apply than the BCT. Both the iTClamp and the BCT were more effective than simple DMP. The iTClamp offers an additional option for managing hard-to-control bleeding in the neck.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Journal of Emergency Medicine
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Thompson E.C.
        • Porter J.M.
        • Fernandez L.G.
        Penetrating neck trauma: an overview of management.
        J Oral Maxillofac Surg. 2002; 60: 918-923
        • Hardaway 3rd, R.M.
        Viet Nam wound analysis.
        J Trauma. 1978; 18: 635-643
        • Chan R.K.
        • Siller-Jackson A.
        • Verrett A.J.
        • Wu J.
        • Hale R.G.
        Ten years of war: a characterization of craniomaxillofacial injuries incurred during operations Enduring Freedom and Iraqi Freedom.
        J Trauma Acute Care Surg. 2012; 73: S453-S458
        • Wade A.L.
        • Dye J.L.
        • Mohrle C.R.
        • Galarneau M.R.
        Head, face, and neck injuries during Operation Iraqi Freedom II: results from the US Navy-Marine Corps Combat Trauma Registry.
        J Trauma. 2007; 63: 836-840
        • Kim J.P.
        • Park J.J.
        • Won S.J.
        • Woo S.H.
        Penetrating carotid artery injuries treated by an urgent endovascular stent technique: report of two cases.
        Chonnam Med J. 2011; 47: 134-137
        • Tisherman S.A.
        • Bokhari F.
        • Collier B.
        • et al.
        Clinical practice guideline: penetrating zone II neck trauma.
        J Trauma. 2008; 64: 1392-1405
        • Kaya K.H.
        • Koc A.K.
        • Uzut M.
        • et al.
        Timely management of penetrating neck trauma: report of three cases.
        J Emerg Trauma Shock. 2013; 6: 289-292
        • Nason R.W.
        • Assuras G.N.
        • Gray P.R.
        • Lipschitz J.
        • Burns C.M.
        Penetrating neck injuries: analysis of experience from a Canadian trauma centre.
        Can J Surg. 2001; 44: 122-126
        • McConnell D.B.
        • Trunkey D.D.
        Management of penetrating trauma to the neck.
        Adv Surg. 1994; 27: 97-127
        • Sperry J.L.
        • Moore E.E.
        • Coimbra R.
        • et al.
        Western Trauma Association critical decisions in trauma: penetrating neck trauma.
        J Trauma Acute Care Surg. 2013; 75: 936-940
        • Weppner J.
        Improved mortality from penetrating neck and maxillofacial trauma using Foley catheter balloon tamponade in combat.
        J Trauma Acute Care Surg. 2013; 75: 220-224
        • Butler Jr., F.K.
        • Holcomb J.B.
        • Giebner S.D.
        • McSwain N.E.
        • Bagian J.
        Tactical combat casualty care 2007: evolving concepts and battlefield experience.
        Mil Med. 2007; 172: 1-19
        • Filips D.
        • Logsetty S.
        • Tan J.
        • Atkinson I.
        • Mottet K.
        The iTClamp controls junctional bleeding in a lethal swine exsanguination model.
        Prehosp Emerg Care. 2013; 17: 526-532
        • Mottet K.
        • Filips D.
        • Logsetty S.
        • Atkinson I.
        Evaluation of the iTClamp 50 in a human cadaver model of severe compressible bleeding.
        J Trauma Acute Care Surg. 2014; 76: 791-797
        • Thompson L.
        Application of the iTClamp in the management of haemorrhage: a case study.
        J Paramed Pract. 2014; 6: 180-181
        • Hudson A.
        • Glazebrook W.
        First UK use of the iTClamp™ haemorrhage control system: case report.
        Trauma. 2014; 16: 214-216
        • St John A.E.
        • Wang X.
        • Lim E.B.
        • Chien D.
        • Stern S.A.
        • White N.J.
        Effects of rapid wound sealing on survival and blood loss in a swine model of lethal junctional arterial hemorrhage.
        J Trauma Acute Care Surg. 2015; 79: 256-262
        • Gilroy D.
        • Lakhoo M.
        • Charalambides D.
        • Demetriades D.
        Control of life-threatening haemorrhage from the neck: a new indication for balloon tamponade.
        Injury. 1992; 23: 557-559
        • Burgess C.A.
        • Dale O.T.
        • Almeyda R.
        • Corbridge R.J.
        An evidence based review of the assessment and management of penetrating neck trauma.
        Clin Otolaryngol. 2012; 37: 44-52
        • Ball C.G.
        • Wyrzykowski A.D.
        • Nicholas J.M.
        • Rozycki G.S.
        • Feliciano D.V.
        A decade's experience with balloon catheter tamponade for the emergency control of hemorrhage.
        J Trauma. 2011; 70: 330-333
        • Carey J.N.
        • Minneti M.
        • Leland H.A.
        • Demetriades D.
        • Talving P.
        Perfused fresh cadavers: method for application to surgical simulation.
        Am J Surg. 2015; 210: 179-187
        • Carey J.N.
        • Rommer E.
        • Sheckter M.
        • et al.
        Simulation of plastic surgery and microvascular procedures using perfused fresh human cadavers.
        J Plast Reconstr Aesthet Surg. 2014; 67: e42-e48
        • Aboud E.
        • Aboud G.
        • Al-Mefty O.
        • et al.
        “Live cadavers” for training in the management of intraoperative aneurysmal rupture.
        J Neurosurg. 2015; 123: 1339-1346
        • Varga S.
        • Smith J.
        • Minneti J.
        • et al.
        Central venous catheterization using a perfused human cadaveric model: application to surgical education.
        J Surg Educ. 2015; 72: 28-32
        • Winer J.L.
        • Kramer D.R.
        • Robison R.A.
        • et al.
        Cerebrospinal fluid reconstitution via a perfusion-based cadaveric model: feasibility study demonstrating surgical simulation of neuroendoscopic procedures.
        J Neurosurg. 2015; 123: 1316-1321
        • Kragh Jr., J.F.
        • Murphy C.
        • Steinbaugh J.
        • et al.
        Prehospital emergency inguinal clamp controls hemorrhage in cadaver model.
        Mil Med. 2013; 178: 799-805
        • Kragh Jr., J.F.
        • Johnson J.E.
        • Henkel C.K.
        • Dubick M.A.
        Technique of axillary use of a Combat Ready Clamp to stop junctional bleeding.
        Am J Emerg Med. 2013; 31: 1274-1276